Specification |
Detail |
Communication |
Link2sc |
Dimensions (H x W x D) |
195 mm x 409 mm x 278 mm |
Display |
7" WVGA (800 x 480) color TFT-LCD with touch screen |
Enclosure Construction |
PC/ABS Blend Plastic |
Light source |
Class 2 laser product, with embedded 650 nm (EPA 0.43 mW) or Class 1 laser product, with embedded 850 nm (ISO), max. 0.55 mW (complies with IEC/EN 60825-1 and to 21 CFR 1040.10 in accordance with Laser Notice No. 50) |
Model |
TU5200 EPA |
Operating Humidity |
80% at 30 °C (non condensing) |
Operating temperature range |
10 - 40 °C |
Power requirements (Hz) |
50/60 Hz |
Power requirements (Voltage): |
100 - 240 VAC |
Range |
EPA: 0 - 700 NTU / FNU / TE/F / FTU; 0 - 100 mg/L; 0 - 175 EBC |
|
ISO: 0 - 1000 NTU / FNU / TE/F / FTU; 0 - 100 mg/L; 0 - 250 EBC |
Regulatory |
EPA |
Repeatability |
<40 NTU: Better than 1% of reading or ±0.002 NTU on Formazin at 25 °C (77 °F), whichever is greater; >40 NTU: Better than 3.5% of reading on Formazin at 25 °C (77 °F) |
Resolution |
0.0001 NTU / FNU / TE/F / FTU / EBC / mg/L |
Response time |
Signal averaging off: 7 seconds; Signal averaging on: 10 seconds (when averaging time is 5 seconds) |
Sample Temperature |
4 - 70 °C |
Source Lamp |
EPA Version - 650 nm Class 2 Laser |
Storage conditions |
-30 - 60 °C |
Stray Light |
<10 mNTU |
Units |
NTU, FNU, TE/F, FTU, EBC; mg/L if calibrated with Degrees calibration curve |
Warranty |
12 months |
Weight |
2.4 kg |
What's included? |
TU5200 benchtop turbidimeter, user manual, Stablcal kit (10, 20 and 600 NTU), 2 sample vials, power supply, dust cover, vial stand |